Almost a third of adult CT scans exceeded dose thresholds in a 254-facility audit

Audit CT dose protocol by protocol; average figures hide the categories running well above threshold.

Design
Evaluation of an automated national quality measure
Population
783,634 adult diagnostic CT examinations at 254 US facilities
Primary outcome
Percentage of examinations above dose or noise thresholds
Effect
30.7% above dose threshold; 0.06% above noise; facility range 1.8–82.4%

This evaluation covers the first 16 months of a US CMS quality measure that automatically checks every adult diagnostic CT against category-specific thresholds for size-adjusted radiation dose and image noise. It included 783,634 examinations across 16 CT categories at 254 facilities, from May 2025 to August 2026.

Overall, 30.7% of examinations exceeded the dose threshold; only 0.06% failed on noise. Facility scores ranged from 1.8% to 82.4% out of range (median 29.5%). Routine abdomen and pelvis, the highest-volume category, had a median of 43.9% out of range, and high-dose head 82.6%. Within the same facility, the gap between its best and worst category was a median of 74 percentage points.

The near-absence of noise failures suggests many protocols run higher dose than image quality requires. Because most variation sat at the level of specific protocol in specific facility, a department-wide average hides where the problem is. The useful step for any CT service, in or outside the US, is to audit size-adjusted dose by protocol category.

  • Audit CT dose by protocol category, not as a single department average
  • Start with routine abdomen and pelvis, the highest-volume and often highest out-of-range category
  • Use size-adjusted dose metrics so larger patients are not flagged unfairly
  • Check image noise alongside dose before lowering a protocol
  • Involve a medical physicist in protocol revision

Why it matters

Most excess dose sits in specific protocols, which makes it fixable once found.

Don't overread it

Exceeding a threshold is a quality flag, not evidence of harm to an individual patient.

The statistics, in plain English

An intraclass correlation of 0.12 means a facility's performance in one CT category tells you little about its performance in another. That is why the authors argue each facility needs category-level results to know which protocols to fix.

Read the rest in the app

You have read your two free briefings this month. The app carries all 27 specialties, every morning, free — and this finding is waiting in it.

QR code to install Daily Doctor
Get Daily Doctor — free

Scan to keep reading on your phone. No account needed to start.

contrastsafetynuclearimaginginterventionalbreastimaging

Tomorrow morning, before your first patient

One edition a day for radiology, written by the desk, every claim tied to its paper. Six minutes.

Get the app — free
Daily Doctor All 27 specialties, every morning. Free.
Get the app